Airborne Particles | Clean Environment | Environmental Monitoring
Vacuum Particle Counter: KKJ-S
It is designed for particle detection in air, vacuum and on-line monitoring environments, and supports particle count evaluation for clean environments and manufacturing processes. This model features multiple alarm modes and remote control via WIFI/LAN. The specific technical configurations and specifications shall be subject to the original catalogue.
Related Products
Test Method: Laser Particle Detection; The specific optical principle shall be confirmed according to the technical description of this model
Manual Reference: GMP is mentioned in the manual; please refer to the full “Performance Features” below for detailed compliance description.
- Support connection to PC data acquisition system for remote control. Real-time test status can be monitored directly; test data can be analyzed and processed on PC and saved as Excel files.
- Technical specifications comply with the requirements of JJG 547-88 issued by the National Metrology Administration.
- Semiconductor laser with service life over 35,000 hours.
- Optional accessories: external printer, dedicated triangular sampling stand, temperature and humidity sensor.
- UCL reports conform to ISO 14644-1, GB 50073-2001, GMP and FS-209E standards.
- Grade alarm function available: triggers alarm for exceeding limits of cleanroom classes (Class 100, 1000, 10000, 100000, 300000, 1000000) and low battery voltage.
Particle Size Range: 0.3, 0.5, 1.0, 3.0, 5.0, 10 μm (6 channels)
Data Storage: 1000 records
Sampling Flow Rate: 2.83 L/min (0.1 ft³), imported sampling pump
Battery: Lithium-ion battery, 7.4 V, 2200 mAh
Counting Mode: Cumulative value, differential value, concentration value
Weight: 0.6 kg
Sampling Period: 1~10000 s
Overall Dimensions: 230×130×45 mm
Delay Counting: 0~99 s
Power Supply & Power Consumption: AC adapter, AC: 100 V~245 V, 50/60 Hz to DC: 7.4 V, 1 A
Self-cleaning Time: ≤10 min
Zero Count: <1 particle / 5 min
Operating Environment: Temperature 10~40 ℃
Repeatability Error: 5% at 2,000,000 particles/ft³
Relative Humidity: 20~90% RH
Sampling Tube / Sampling Probe (verified according to flow rate and interface) | Filter Assembly (according to application configuration) | Standard instruments for calibration and count verification (not included as standard accessories)
Items marked in green are recommended supporting list instead of standard built-in accessories; please confirm according to actual interface, material and packing list.